Abstract

We have designed, constructed, and tested a unique cold finger suitable for angle resolved photoemission spectroscopy. This design is based on in situ helium reliquification and utilizes pulse tube cryocooler. The pulse tube can be removed for baking without breaking Ultra High Vacuum (UHV). This design also allows the use of non-UHV heater that can be replaced without the need to vent the system. The cold finger has minimal vibration, operates over a temperature range of 1.7 K–400 K, and has no measurable residual magnetization. In continuous mode, it can maintain a sample temperature of 2.6 K, while in single shot mode (by pumping on liquid helium), it can reach temperatures down to 1.8 K for a period of several hours.

DOE Contract Number(s)

AC02-07CH11358

Language

en

Publisher

Iowa State University Digital Repository, Ames IA (United States)